| SPECIFICATIONS |
| General |
| Compatible Sensor Types | All models of PyroMiniBus and PyroBus sensors; -BB and -BRT models of PyroMini and FibreMini sensors. Up to 6 sensors per PM180. |
| Display | 2.83” (72 mm) resistive touch TFT, 320 x 240 pixels, backlit |
| Supply Voltage | 24 V DC (min. 10 V DC / max. 30 V DC) |
| Maximum Current Draw | 100 mA |
| Configurable Parameters (global) | Temperature units, date and time, data logging, graph channels, alarm logging |
| Configurable Parameters (per channel) | Signal processing, emissivity setting, reflected energy compensation, alarms, Modbus address |
| Alarm Configuration | 12 alarms (2 per sensor) with adjustable level,individually configurable as HI or LO. |
| Temperature Units | °C or °F selectable |
| Temperature Resolution | 0.1° |
| Signal Processing | Averaging with configurable period |
| Display Sample Period | 120 ms per sensor (720 ms in total for 6 sensors) |
| Data Logging |
| Logging Interval | 1 to 86,400 seconds (1 day) |
| MicroSD Card | Max. capacity: 32 GB (not included - stores years of logged data) |
| Internal Clock Battery | 1 x BR 1225 3V (not included) |
| Variables Logged | Target temperature, sensing head temperature, alarm events |
| File Format | .csv (can be imported to Excel) |
| Configurable Parameters | Sample period, number of samples, scheduled start date and time |
| Mechanical |
| Construction | Die Cast Aluminium |
| Electrical Connections | Removable screw terminals, 28 AWG to 18 AWG |
| Dimensions | 98 (w) x 64 (h) x 36 (d) mm excluding cable glands |
| Weight | 280 g |
| Environmental |
| Environmental Rating | IP65 |
| Ambient Temperature | 0°C to 60°C |
| Relative Humidity | Maximum 95%, non-condensing |
| Environmental |
| RoHS Compliant | Yes |
| Electromagnetic Compatibility | EN61326-1, EN61326-2-3 (Electrical Equipment for Measurement, Control and Laboratory Use - EMC Requirements - Industrial) |
| Language |
| Languages | Selectable: English, Chinese (Simplified) or Japanese |