Specifications |
ANALOG SECTION |
Excitation voltage | DC3V±10% Output current: within 35mA |
Signal input range | -3.0 - +3.0mV/V |
Equiv. input calibration range | +0.5 - +3.0mV/V |
Zero adjustment range | -2.0 - +2.0mV/V |
Gain adjustment range | +0.5 - +3.0mV/V |
Accuracy | Non-linearity: within 0.02%/FS (at 3.0mV/V input) Zero drift: within 0.3μV/℃ RTI Gain drift: within 5ppm/℃ |
A/D converter | Speed: 80 times/sec. Resolution: 24bit(binary)Approx. 1/30000 at 3.0mV/V input |
DISPLAY SECTION |
Display unit | 128×64 dot black and white LCD |
Measured value | 5digits: -99999 to +99999 Sign: Character height: 14mm |
Status display | Status display1……R(Recording) / A(AC adapter on use) / U(USB in connection) / N(NOV RAM reading) / B(Backup battery abnormal) Status display2……HI/OK/LO/PEAK/BOTTOM/HOLD |
SETTING SECTION |
Setting items | ・Tare*/ equivalent input calibration/ span calibration/ zero calibration(one-touch zero)/ weight unit setting*/ channel setting/ HI setting/ LO setting/ machine ID setting*/ setting value lock ・Unit/ digital filter/ MD(period, range)/ZT(period, range)/ near zero/ digital zero/ minimum scale ・Data/ buzzer/ backlight/ auto off/ version display/ memory check/ initialization ・Hold: mode/ range/ start level/ end level/ monitoring time ・Graph: time axis*/ mode*/ trigger edge*/ trigger level*/ pre-trigger* ・Operation mode/ HI/LO comparison mode/ measuring mode*/ recording mode*/ real-time output* * not usable at Simple Operation Mode |
RECORDER |
Recording function | ・Record when [REC] is pressed ・Record when stability is detected ・Record hold value at hold-release time ・Interval recording(records data at every fixed interval) ・Records graph data(records data displayed in graph) |
Recording media | Internal memory |
Recording method | Text form of CSV format |
Recording data | ID, number of recordings, recording data, measured channels, measured values, unit |
Number of recording data | 20,000 data |
HOLD SECTION |
| Hold Mode: Peak Hold/ Valley Hold/ Peak-to-Peak Hold/ Sample Hold Data monitoring range: All ranges/ level/ level + time |
MEASURING MODE |
| Load measuring/ counting |
INTERFACE |
| USB interface(Communication standard: Compliant with USB Ver.1.1 Communication speed: Full speed Class: Mass-storage and virtual COM port OS: Windows XP (SP3),Windows Vista (SP1)) ・File operation The F490A is recognized as a drive, which enables file reference and copying. ・Virtual COM port Set values can be read and written by terminal software. ・Real-time output Measured data are output in succession on USB. |
GENERAL SPECIFICATIONS |
Internal power source | AA alkaline batteries or nickel metal hydride batteries (4) |
External power source | Dedicated AC adapter (for 100V AC) |
Current consumption | When a 120Ω sensor is connected: Approx. 60mA (back light off) When a 120Ω sensor is connected: Approx. 70mA (back light on) |
Backup power source | Set values and record data are maintained by the lithium battery (warranty period 5 years or more) |
Continuous use time | When a 350Ω sensor is connected: Approx. 30 hours (back light off) When a 120Ω sensor is connected: Approx. 12 hours (back light off) |
Operating conditions | Temperature: -10℃ to +40℃ Humidity: 80%RH or lower(non-condensing) |
External dimensions | 86(W)×132(H)×30(D)mm(not including projections) |
Weight | Approx. 290g (including the battery weight approximately 95g) |
ACCESORIES |
Sensor connector | 1 |
AA alkaline battery | 4 |
CD-ROM | 1 |
Operation manual | 1 |